Seven Common Faults of Elliptic Gear Flowmeters and Their Treatments

Oval gear flow meter volumetric flow meter, also known as fixed-displacement flow meter, is the most accurate type of flow meter in all flow meters. It is insensitive to flow distortion, can measure non-Newtonian flow, and can reflect true volumetric flow in a miscible flow measurement.

First, the oval gear does not rotate

Reasons: 1. There are debris in the pipeline;

2, the filter is damaged, debris into the table, causing the gear to be stuck;

3, due to excessive flow, bearing wear, resulting in elliptic gear collision guilt, causing gear stuck.

Treatment method: 1. Clean the flowmeter and pipeline;

2, repair filters;

3, replace the bearings to ensure that the gap can run flexibly.

Second, the oval gear rotation, the pointer does not move or when stopped

Causes: 1. Rotating fasteners sliding, damaged or inflexible;

2, magnetic coupling sliding;

3, word wheel parts are stuck.

Treatment methods: 1, re-fasten the fasteners, remove inflexible places;

2, check the magnetic coupling is flexible and magnetic strength;

3, troubleshooting the word wheel components.

Third, the flowmeter negative error

Causes: 1. Bypass valve leakage;

2. The internal wall or bearing of the elliptical gear measurement is worn or damaged;

3, the flow rate is lower than the lower limit;

4, magnetic coupling sliding or gear system periodic obstacles;

5, the actual fluid viscosity is less than the detection fluid viscosity.

Treatment method: 1, check the bypass valve;

2, replace the bearing, check the gear, adjust the gap;

3, increase the flow;

4, repair or replace the magnetic coupling or gear system;

5. Select a test fluid with a viscosity similar to the actual liquid to check the flowmeter or make a viscosity correction.

Fourth, the positive error of the flowmeter

Causes: 1. The flow has large pulsations or contains gas;

2, the actual liquid viscosity is greater than the viscosity of the test liquid;

3. When the inlet valve is closed, the medium is not filled with the metering chamber.

Treatment methods: 1. Reduce flow pulsation or increase gas separators;

2. Select a test fluid with a similar viscosity to the actual liquid to test the flowmeter or make a viscosity correction;

3, use the flow meter outlet valve to regulate the media flow.

Five, secondary table indicating value more than one table indicating value

This is caused by the fact that the failure is caused by the vibration of the transmitting blade, which can be solved by appropriately adjusting the pressure of the damping plate to the blade.

Sixth, secondary table indicator is less than the value of a table indicator

Causes: 1. Loosening screws at the transmitting blade;

2, sender out of step.

Treatment methods: 1, fastening the locking screw;

2. Adjust the gap between the transmitting blade and the transmitter.

Seven, the second table does not indicate

Reasons: 1. Open circuit or short circuit;

2. The transmitter is damaged.

Treatment method: 1, carefully check the line;

2, replace the sender.

Double Wire Fence

QYM-Double Wire Fence

It is a special design fence, which have a double horizontal wire, in this way the fence can be much more stronge, and beautifull looking,The suitable post is made from RHS pipe, connected by steel clamp.China Double Wire Fence,Double Wire Mesh Fence manufacturers, welcome Double Sided Wire Fence,Double Wire Welded Mesh Fence purchasers from worldwide to visit our site.

double mesh fence 1-1


Double Wire Fence,Double Wire Mesh Fence,Double Sided Wire Fence,Double Wire Welded Mesh Fence

Hebei Qianye Metal Product Co., Ltd. , https://www.qymetalfence.com

Posted on